| RADIOGRAPHIC SURVEY METER 3 RANGE, 0-1000 MR/HR 
Features : 
  | 
▮ Specification
| Ranges - Dual Scale | 0 - 1,000, 0 - 100, 0 - 10 mR/hr 0 - 10,000; 0 - 1,000; 0 - 100 uSv/hr | 
| Efficiency | within 10% of full scale at standard temperature over operating range | 
| Energy response | Gamma and X-Ray from 80-15000keV, +/- 20% | 
| Saturation | In excess of 1000R/HR When calibrated and properly maintained | 
| Environmental effects | Humidity limits : 0-99% NON-Condensing | 
| Detector | Halogen quenched, energy compensated geiger mueller tube Effective length and diameter : 0.625 inch×0.194 inch wall thickness : 90mg/cm | 
| Controls | 5 Position rotary range switch : off; X-100, X-10, X-1 and battery / High voltage test | 
| Calibration controls | Individual Potentiometers for adjustment of each Range | 
| Calibration Precision | Within 10% when calibrated with Cs-137 | 
| Response time | 90% of final reading within 10 seconds | 
| Geotropisim | within 2% of full scale in any orientation | 
| Dimensions | 4.4(W) × 7.75(L) × 6.25(H) inches | 
| Weight | 1.73kg(including batteries) | 

 
 
 
 
 
 
 
 
 